PEM (Center)
takaaki.yokozawa - 12:53 Thursday 04 May 2023 (25114) Print this report
Shaker injection at OMC chamber
[YokozaWashimi]

We performed the shaker injection for OMC chamber.
We can see some excess in several frequencies.

1. around 80 Hz
2. around 290 Hz
3. around 560 Hz
4. around 800 Hz
The analysis results will be appeared later.

Unfortunately, we can only one position and we could not check around 800 Hz in detail.
So if there are next chance, we will try to do it.
Images attached to this report
Comments to this report:
tatsuki.washimi - 17:39 Thursday 04 May 2023 (25115) Print this report

The location we shook today was on the shoulder of the -Y side flange because it was easy to fix the shaker around the beam-hight.

Shaking tests for other locations were not performed due to the instability of the IFO.

Images attached to this comment
tatsuki.washimi - 17:54 Thursday 04 May 2023 (25116) Print this report

This is result of the broad-band injection for 3 amplidtudes

Images attached to this comment
tatsuki.washimi - 8:44 Friday 05 May 2023 (25117) Print this report

I analyzed today's data and estimated the noise projection.

The formula used is the same as written here.

It is not sure if this analysis is appropriate, but if it is correct, the noise contribution from this location is much lower than the sensitivity.

Images attached to this comment
Non-image files attached to this comment
takaaki.yokozawa - 6:36 Saturday 06 May 2023 (25120) Print this report
What accelerometer did you use for your analysis?
In case you used nearest accelerometer from shaker, there are possibility of under estimation.
It may depend on the noise path, could you try similar analysis using other accelerometer?
Or shaker test again using another shaker point.
tatsuki.washimi - 14:29 Saturday 06 May 2023 (25121) Print this report

The ACC used in this analysis was "ACC3", near the shaker.