When:
2nd/Feb./2023
Workers:
N. Kimura
H. Yasui
M. Takahashi
Task:
We performed a vacuum leak test for OMMT and OMC fter leak repair (https://klog.icrr.u-tokyo.ac.jp/osl/?r=23775) using a helium leak detector and helium gas.
The tests were performed on all flange connections except the top flange of the OMMT and OMC and the area on the ASC side of the OMC.
The test was conducted with a test sensitivity of ~1 x 10^-12 Pam^3/s, confirming that there was no leakage > ~1 x 10^-11 Pam^3/s .
Helium background during the test ranged from 8.5x 10^-10 Pam^3/s to 1 x 10^-10 Pam^3/s.
Leakage testing of the flanges in the area on the ASC side is scheduled for 3rd/Feb..