[Yokozawa, Yuzu, Yamazaki, Murakoshi, Obayashi]
The motivation for this study is to evaluate the stationarity of environmental monitors.
When the candidate of GW signal is detected, the GW signal can be validated by using the well-characterized environmental monitors.
As a first step, we analyzed the data of the accelerometers (K1:PEM-ACC_EXA_TABLE_TX_Z_OUT_DQ, K1:PEM-ACC_EXA_TABLE_RX_Z_OUT_DQ).
We plotted an averaged ASD with percentiles for each 10-minute interval for one day of these data on 2022/4/7 17:00 - 2022/4/8 17:00 (during O3GK). See the attached images.
We found that a line-like feature around 40 Hz may be non-stationary.
Future work
- By using one week data, check behavior and trends.
- Same work at the other sensors.