Date: 2022/8/24
With Shingo Fujii
We performed a calibration process for YPcal.
quick result
Alpha values are voltage ratio between ISs in XPcal and GSK(new name of this integration sphere is WSK). [Main-value Uncertainty-value]
alpha_RxPDpGSK = [0.80513707 0.0008152 ]
alpha_TxPD1pGSK = [4.50914710e+00 9.30131312e-04]
alpha_TxPD2pGSK = [1.60451441e+00 9.68060082e-04]
e values are optical efficiency values. [Main-value Uncertainty-value]
e_1 = [9.69337449e-01 2.60179127e-04]
e_2 = [1.00473157 0.00164785] (Incidence angle dependent effects not yet included)
A detailed report will be written on CAL wiki: link