VAC (Tube Y)takaaki.yokozawa - 18:16 Friday 10 June 2022 (21100)
Print this reportComment to Increasing the vacuum level of Yarm (21071)I also checked the vacuum level.
(Yesterday) -> (Today) BS-IYC : 2.3e-4 Pa -> 2.2e-4 Pa IYC-IYA : 2.2e-4 Pa -> 2.1e-4 Pa IYA-GV : 1.8e-4 Pa -> 1.8e-4 Pa Y2 : 3.8e-5 Pa -> 3.7e-4 Pa Y10 : 2.0e-5 Pa -> 2.1e-5 Pa Y15 : 1.6e-6 Pa -> 1.9e-6 Pa Y21 : 2.8e-6 Pa -> 2.9e-6 Pa Y27 : 2.5e-5 Pa -> 2.4e-5 Pa
And finally I found the gate value at Y-10 was CLOSED. I found some klog(https://klog.icrr.u-tokyo.ac.jp/osl/?r=18535), previously troubled. From that time (October 2021), this gate valve was closed, so I opened gate valve again and I will monitor the vacuum level this weekend. After opening the gate valve, the value of cc-10 became 1.5e-5 Pa
I tried the check the exact time to increase the vacuum level First increasing occurred at 2:00 a.m. 8th June (Early Wednesday)