VIS (PR2)takaaki.yokozawa - 14:21 Monday 16 May 2022 (20741)
Print this reportCharacterization of PR2 oplev[YokozaWashimi]
We performed several characterization for PR2 oplevs. Before this work, we confirmed the IR alignment by locking the Xarm and I asked to Ushiba-san to turn off the damping using TM oplevs.
1. Changing the position of TILT QPD Because there are so small space between beam splitter for length sensing QPD, we slightly changed the position of TILT QPD. Now there are some space (We can place the beam profiler, so we may install the cover and filters in future)
2. Setting the length sensing QPD With adding the oscillation to K1:VIS-PR2_TM_TEST_P_EXC (0.43 Hz and 30 cnt) Fig.1. showed the spectrum after setting the length sensing QPD.
3. Measured the calibration factor TM TILT VER : 1.25 [cnt/mm] (Fig.2.) TM TILT HOR : -1.29 [cnt/mm] (Fig.3.) TM LEN HOR : -4.72 [cnt.mm] (Fig.4.)
4. Measured the distance of each optics
d_vb = 115 mm d_bt = 80 mm d_bl = 65 mm d_lm1 = 100 mm d_m1m2 = 215 mm d_m2l = 60 mm
5. Measured the beam power Input : 0.658 mW TILT QPD : 0.061 mW (15551 cnt with high gain mode) LEN QPD : 0.120 mW (23729 cnt with high gain made)
6. Measured the beam profile (pdf1 and pdf2) TILT width major 2.343 mm TILT width minor 2.154 mm LEN width major 0.5264 mm LEN width minor 0.3998 mm The shape of beam profile of TILT QPD is not so clean, why??
takaaki.yokozawa - 14:32 Monday 16 May 2022 (20742)
Print this reportWith believing those values (790 mm and 49 deg) in Fig.1 I calculated the calibration factor
takafumi.ushiba - 18:27 Wednesday 22 June 2022 (21240)
Print this report
I implemented the calibration factors for PR2 OpLev to medm today. Related to this work, I changed several things for compensate the calibration factors effect.