Summary: After the calibration of the F0 LVDT, I wrote the appropriate calibration factor and offset in the medm screen.
- The slope from the plot is 1348.1 cnt/mm, which becomes 0.7418 um/cnt for the medm screen.
- The LVDT output at the 3D CAD nominal position of the keystone (16.5 mm) is -1461.438 cnt, therefore, the offset to write in the medm screen is 1461.438 cnt.
- After the change I accepted the new values in the SDF.