fabian.arellano - 12:34 Friday 15 October 2021 (18586) Print this report
Comment to Geophone calibration, noise measurement (7059)

I calculated the theoretical sensitivity of the geophones used in SR3:

  • Geophones names: 1, 2 and 3.
  • The numerical data of the geophones sensitivities is not available, so I took it from the plots using WebPlotDigitizer.
  • Below 500 mHz, the calculated and measured values seem to coincide.
  • As far as I remember being told, these geophones have been very noisy after installation on the IP table. I need to measure the IP free swing displacement ASD on a day with low micro-seismic in order to compare.