fabian.arellano - 17:33 Wednesday 13 October 2021 (18558) Print this report
Comment to Geophone calibration, noise measurement (7059)

I calculated the theoretical sensitivity of the geophones used in SR2:

  • Geophones names: 4, 5 and 6.
  • The measured and calculated sensitivities are shown in the first screenshot. The noise is higher than expected.
  • The numerical data of the geophones sensitivities is not available, so I took them from the plots using WebPlotDigitizer.
  • At 200 mHz, the measured values are higher than expected by the following factors:
    • GEO4: 2.7
    • GEO5: 3.7
    • GEO6: 4.7
  • The plot also shows the IP-L free swing displacement measured with the geophones on a day with good weather. It was measured by Terrence and Lucia on the 20th of July 2020 (klog 14790 maybe).
  • The second screen shot shows the noise budget of GEO4 as an example. The dominant contribution is the pre-amplfier noise.
  • The third screenshot shows a comparison of two different pre-amplfier calculations and a measurement. The three are consistent.
  • I used this reference: Blom, M. (2015). Seismic Attenuation for Advanced Virgo: Vibration Isolation for the External Injection Bench [Phdthesis]. Vrije U., Amsterdam, page 152.

I will add measurements of the IP free swing spectra measured by individual geophones when the micro-seismic is low.