Reports 1-1 of 1 Clear search Modify search
VIS (General)
satoru.ikeda - 14:18 Friday 01 October 2021 (18418) Print this report
Comment to Routine confirmation of DGS maintenance day (17831)

Routine confirmation of DGS maintenance day (10/01)

Routine check for DGS maintenance day (10/01) was performed.
Two problems were found.
The limit switch of BS_F0YAW was in contact with both L and R.

The LAN of ETMY Pico and the temperature control device could not be connected.
 => I changed the power supply of the network SW to PoE on 9/27, but it recurred after that.

Stepper motor
 [Checked]
  Checked if the power can be controlled by BIO.
  Checked if Script can be started.
  Checked that the limit switch installed is green.
 [Check result]
  The limit switch of BS_F0YAW was abnormal because both L and R were in contact.
  => Confirmation required.
  No other problems.

Release Sensor
 [Confirmation]
 Confirmed that Release Sensor is 0. 1: Seated
 [Check result]
 ETMY, ITMX, ITMY : No problem
 ETMX_F1_1 : abnormal, no other problems (known)

Picomotor
 [Checked]
  PingTest is performed.
 [Result]
 NG:
  ETMY
   => Unable to access from LAN after 9/27 11:38.
    Will check on Monday.
  SR2_IM,SR2_STM
   => Both have been rebooted and recovered.

  MCI,STM1,IMMT1,IMMT2:NG
  AS_WFS,POP_POM,POS_POM,REFL_WFS,POM1: NG
   => Confirmation only, no recovery work done.

 OK:
  ETMX,ITMX,ITMY
  BS(IM,BF),SRM(IM,BF),SR2(BF),SR3(IM,BF)
  PRM(IM,BF),PR2(IM,BF),PR3(IM,BF)
  MCF(MCO,MCI),MCE
  OMMT1,OMMT2,OSTM

  PCAL(EX1,EX2,EY1,EY2):Not confirmed because it is ON only when used.

Temperature Control
 [Checked]
 Check if the temperature acquisition is working properly.

 [Check result]
 ETMY:NG
   => Unable to access from LAN after 9/27 11:38.
    Will check on Monday.
  Other than the above, everything was OK.
 

Images attached to this comment
Non-image files attached to this comment