takahiro.yamamoto - 20:45 Wednesday 07 July 2021 (17430) Print this report
Comment to Trial of GreenX reflection from Xend Day3 PM: GrX returns (17428)
Attachment shows the RFDC signal when the reflection light is absent and presence.
When reflection light is absent, this channel is limited by the circuit noise which is slightly larger than the ADC noise.

In the case reflection light is presence, K1:ALS-X_ARM_REFL_IN1 which is raw ADC count shows ~6.4ct = 3.9mV.
Universal whitening filter is now set as 12dB gain at DC. So the PD output should be 1mV (at differential) and 500uV (at single-end).

Considering 100Ohm as an impedance and 0.3A/W as a PD efficiency, the laser power at RFDC is ~16uW.
Images attached to this comment