Miyo, Ushiba, T. Yamamoto, Akutsu: Cont'd from 17140, 17418, 17419
Abstract
The green beam came back from ETMX to GrRFPD on the POP table.
Details
First we misaligned ITMX so that the reflection beam from ITMX went out of the GrRFPD. Immediately after Miyo-kun started sweeping ETMX, we found a periodic and repeatable responses in the GrRFPD (see Figs 1, 2 and 3 from Miyo-kun); as reported in 17418, 17419 by Ushiba-kun and Yamamoto-kun, the signal read of the GrRFPD (K1:ALS-X_ARM_REFL_OUT16
) should be improved in S/N. Then we went to the Xend and I tweaked the oplev QPD stages to zero for memorizing the "best" alignment.
Post analysis
So, it seems K1:ALS-X_ARM_REFL_OUT16
provides "0.004" when the relfection from ETMX (I believe so...) enters GrRFPD. In watt, what is this 0.004?? Unfortunately I could not see this returning beam with my eyeballs. There are several non-empty filters in the bank (see 17418) and so we need to understand what are them for proper calibration.
Memo: the conversion efficiency of the PD is ~0.3A/W (S3399 datasheet; about the modification of the RFPD from the usual IR RFPD is found in 5735)