Reports of 29864
VAC (Valves & Pumps)
nobuhiro.kimura - 7:56 Saturday 11 January 2025 (32312) Print this report
Comment to GVbsx and GVbsy have been closed (32286)

 

Results of the partial pressure measurement at 7:43 on Jan. 11 are as follows; 

Place

 BS

BS

IXC

IYC

Date

1/07

1/11

1/11

1/11

Time

 10:45

 7:43

 7:39

 7:44

T.P. (Pa)

 1.28 x 10^-5

 1.20 x 10^-6

 4.37 x 10^-6

 4.62 x 10^-6

H2 (Pa)

 1.42 x 10^-5

 1.52 x 10^-6

 1.65 x 10^-6

 2.01 x 10^-6

H2O (Pa)

 1.57 x 10^-6

 1.49 x 10^-6

 1.27 x 10^-6

 1.22 x 10^-6

N2 (Pa)

 6.18 x 10^-7

 5.05 x 10^-7

 6.22 x 10^-7

 6.04 x 10^-7

O2 (Pa)

 3.04 x 10^-7

 3.10 x 10^-7

 2.21 x 10^-7

 2.03 x 10^-7

 

The partial pressure of the residual gas components were recovered to the level before the interlock device operated and the gate valves closed.

 

VAC (Valves & Pumps)
nobuhiro.kimura - 2:47 Saturday 11 January 2025 (32306) Print this report
Comment to GVbsx and GVbsy have been closed (32286)

[Kimura and Yasui]

 BS CC-10 pressure reached 7.5 x 10^-6 (Pa) at 16:10. 

The threshold of the CC-10 contact output was then set to (7.5 + 0.5) x 10^-6 (Pa) of 8.0 x 10^-6 (Pa). 

After setting, the interlock device was enabled.

 

VIS (EX)
takafumi.ushiba - 18:05 Friday 10 January 2025 (32311) Print this report
ETMX BF yaw setpoint was changed

I changed the ETMX BF yaw setpoint to offload the MNY actuator outputs.
The value was changed from -1700 to 200.

VIS (EX)
takafumi.ushiba - 18:00 Friday 10 January 2025 (32310) Print this report
ETMX MN OpLev seems clipped

ETMX OpLev SUM is now about 100 cnts while it was ~10000 cnts before cooling, so I started to investigation.
Figure 1 shows the MN OpLev SUM when changing the MN Optic align values at PAY_FLOAT state.
When MN Optic align values are -1200 cnts from the aligned position, MN OpLev SUM is more than 9000, which implies that MN OpLev beam is clipped somewhere.

Since the cooling would change the relative angle between TM and MN, this phenomena can happen even if we kept the good alignment of ETMX.
What we can do depends on the situation:
1. If the clipping happens on the BS of the detection port, we can recover it by realigning the optics on the reciever side.
2. If the clipping happens inside the cryostat, we need to move injection beam to recover the clipping.

Anyway, ETMX MN OpLev damping performance getting worse and it is dificult to lock the interferometer stably in this condition, so we need to fix the probem.

Images attached to this report
VIS (EX)
kenta.tanaka - 17:33 Friday 10 January 2025 (32308) Print this report
Turn on 3-stage whitening filters for ETMX TM HPCD VMON

Uchiyama, Tanaka

Ushiba-san asked us to install an oscillation stopper (発振止め) in Fig.1 between the VMON port on TM HPCD and the whitening charsis and to turn on the 3-stage whitening filters.

We installed it (fig.2) and turned on them . Also, we engaged the dewhitening filters in the TM_COILDRV_VMON filterbank (fig.3).

Images attached to this report
IOO (OMC)
kenta.tanaka - 17:20 Friday 10 January 2025 (32304) Print this report
Replacement of the beacon summing TM HPCD was failed

Uchiyama, Tanaka

We tried to replace the TM HPCD to the one for beacon summing (s2315283) at EX.

We requested the ETMX guardian to go to the SAFE state before the replacement. We turned off the old coil driver after ETMX reached the SAFE state. Then, we replaced the coil driver to the one for beacon.

After that, we restored ETMX to the PAY_FLOAT state, and we measured the TFs from ETMX_TM_COILOUTF_H{1,2}_IN2 to ETMX_TM_WIT_P and the TFs from ETMX_TM_COILOUTF_H{3,4}_IN2 to ETMX_TM_WIT_Y at 10 Hz. Fig.1,2,3,and 4 show the results of H1, H2, H3, and H4, respectively.

In the old coil driver setup, the summing adaptor has 2.7dB loss at the input of the driver. So we compensated it with the gain in FM10 of COILOUTF. On the other hand, in the new coil driver setup, there were no loss according to the TF measurement results by the AEL group (JGWdoc). Therefore, the gain of the TF (brown) when we used the new driver and we turned off FM10 of COILOUTF should be the same as the previous results (blue). In this sense, H2 result seem to be fine but the others seem to be ~6dB lower than the ones in the previous setup.

We suspected that the previous health check was performed last June, that is, the payloads were not cooled and the gain became lower by cooling. So we replaced the new driver to the old one. And we measured the TF. The red line in the figures are shown the results. The reds seem to be the same as (or slightly larger than) the blues. Therefore, the cooling seems to be not the cause of -6dB.

We left the old coil driver in the ETMX rack and we brought back the new one. I asked AEL group to investigate the cause.

Images attached to this report
CAL (General)
hirotaka.yuzurihara - 16:37 Friday 10 January 2025 (32303) Print this report
TCam photo session 20250110

I took the TCam photos for four mirrors at 14:42~14:52 after the initial alignment. I diagnosed the analysis results and it's unnecessary to update the reference values.
The mirror appearances were same as usual and healthy.

MIF (General)
hirotaka.yuzurihara - 16:09 Friday 10 January 2025 (32305) Print this report
Comment to Full initial alignment 2025/01/10 Afternoon (32302)

I edited the medm screen for the initial alignment, to remove the misunderstanding.

Images attached to this comment
MIF (General)
hirotaka.yuzurihara - 15:03 Friday 10 January 2025 (32302) Print this report
Full initial alignment 2025/01/10 Afternoon

[Yokozawa, Yuzurihara]
We performed the full initial alignment (Xarm, Yarm, OMC and PRMI) in 14:08~14:40. During Yarm alignment, I changed the setpoint of BS in pitch: 10.9 -> 12.9.

Comments to this report:
hirotaka.yuzurihara - 16:09 Friday 10 January 2025 (32305) Print this report

I edited the medm screen for the initial alignment, to remove the misunderstanding.

Images attached to this comment
VAC (Valves & Pumps)
nobuhiro.kimura - 14:54 Friday 10 January 2025 (32301) Print this report
Comment to GVbsx and GVbsy have been closed (32286)

[kimura and Yasu]

 We turned off two TMPs and two dry-pumps at 14:50.

 

VAC (Valves & Pumps)
nobuhiro.kimura - 14:12 Friday 10 January 2025 (32300) Print this report
Comment to GVbsx and GVbsy have been closed (32286)

[Kimura and Yasui] 

GVbsx and GVbsy were opened in the following steps.

2Step 1: 

 13:45 Disabled the enable/disable switch in the GVbsx and GVbsy interlock device.

Step 2

13:48 Opened GVbsy and closed the GVs of the TMP.
BS-CC-10 pressure was 1.4 x 10^-5 Pa.

Step 3

13:50 Opened GVbsx and closed the GVs of the TMP.
BS-CC-10 pressure was 1.4 x 10^-5 Pa.

Results of the partial pressure measurement are as follows;

 

BS

IXC

IYC

Time

14:03

14:09

14:05

T.P. (Pa)

1.45 x 10^-5

 4.15 x 10^-6

5.73 x 10^-6

H2 (Pa)

2.00 x 10^-6

 1.68 x 10^-6

2.27 x 10^-6

H2O (Pa)

2.09 x 10^-6

 1.16 x 10^-6

1.62 x 10^-6

N2 (Pa)

5.54 x 10^-7

 5.61 x 10^-7

7.03 x 10^-7

O2 (Pa)

3.54 x 10^-7

 2.07 x 10^-7

2.48 x 10^-7

VAC (Valves & Pumps)
shinji.miyoki - 13:49 Friday 10 January 2025 (32299) Print this report
Comment to GVbsx and GVbsy have been closed (32286)

Because TMPs will be off, I recovered the target temp for the precision cooler from 19 to 20 and turned off two Sunrise heaters around 14:00.

VIS (General)
takafumi.ushiba - 11:44 Friday 10 January 2025 (32298) Print this report
Do not forget to load filter banks after changing filters

I found several filter banks (ITMX_NBDAMP_V1, ITMY_NBDAMP_R1, and ITMY_MNOLDAMP_P) not loaded after modifying the filter parameters.
This time, I loaded all of them because these changes are just gains, notch frequencies, and filter name, and I could confirm new filter seems fine.
However, if it is a complex modification, it would take time to confirm loading new filters are good or not.
So, please be careful not to forget to load the filter banks after changing the filters.

VIS (IX)
takafumi.ushiba - 11:36 Friday 10 January 2025 (32297) Print this report
ITMX NBDAMP V1 filter was not loaed

I noticed ITMX NBDAMP V1 filter was designed but not loaded, so I load filter bank and measure the OLTF.
Figure 1 shows the result, and there seems no significant difference between reference (before load) and current one.
So, the lop seems fine.

Images attached to this report
VAC (General)
takaaki.yokozawa - 9:53 Friday 10 January 2025 (32296) Print this report
Changed the sampling rate for the vacuum monitor
I changed the sampling rate of the vacuum monitor from every 5 min to about 25 sec.
(But from the crontab of k1script1, the sampling rate would be about 1min)
VIS (IFI)
Kohei Mitsuhashi - 4:40 Friday 10 January 2025 (32293) Print this report
Measurement of a mirror's spectrum with changing clamp number, clamp position, and type of clamp
with Washimi-san, Tomaru-san, Takahashi-san (online), Ikeda-san

Conclusion


・The spectral structure between 80 and 200 Hz of the mirror on mount similar to IFI STM 2's mount can be changed depending on how the clamps are fixed and how many clamps we use to it.


Details


Introduction


We measured the displacement of a mirror mounted on a structure similar to the IFI STM 2 mirror mount under various experimental conditions. The experimental conditions were as follows:
1. Relationship between the tightening levels of set screws and the mirror’s spectrum
2. Relationship between number and type of clamps and the mirror’s spectrum
3. Relationship between clamps position and the mirror’s spectrum
4. Relationship between tapping locations and the mirror’s spectrum
Displacement sensor's resolution is 0.15 nm.

1. Relationship between the tightening levels of set screws and the mirror’s spectrum


Using the same setup as before (pic1), we varied a tightening levels of a set screws on side of the mirror mount (pic2) and measured the displacement spectrum of the mirror. However, it was difficult to quantify the tightening levels, as the rotation angle of a hex wrench was not visually discernible, and an adjustment relied on tactile feedback. The results (pic2) showed no significant changes, suggesting a minimal relationship between the tightening levels and the spectrum.

2. Relationship between number and type of clamps and the mirror’s spectrum


In the previous Klog report, we used only one clamp, which differed from those used in actual vacuum chambers named Neji clamp (pic3). For this experiment, we included clamps believed to be used in the IFI chamber referred to as L clamp (pic4), examining the relationship between their number, position, and the spectrum.
The experiment was conducted with no clamps, one clamp, two clamps, and three clamps. The clamp types included Neji clamps and those presumed to be used in the IFI chamber. Details of the setups are shown in pic5, and the results in pic6.
The results indicated that the spectral structure in the 80–200 Hz band changed depending on the number of clamps. However, no significant changes were observed below 200 Hz when comparing two and three clamps, whereas changes were observed above 200 Hz.

3. Relationship between clamps position and the mirror’s spectrum


We investigated the impact of clamp positions on the spectral structure and evaluated the reproducibility of these results.
Clamps are circled in red in pictures.
- Single clamp
The setup for varying clamp positions with one clamp is shown in pic7, and the results in pic8. The results confirmed a relationship between clamp positions and the spectral structure in the 100–200 Hz range. Additionally, comparing the results of position 1 and position 1again (clamped again at the same position) demonstrated reproducibility.
- Two or three clamps
The setups for varying clamp positions with two or three clamps are shown in pic9, and the results in pic10. No changes in the spectral structure were observed in the 100–200 Hz range. However, changes were observed above 200 Hz.

4. Relationship between tapping locations and the mirror’s spectrum


To identify which resonance peaks, correspond to specific parts of the structure, we measured the spectrum while tapping different locations. The tapping locations and their corresponding results are attached.
The findings from this experiment are summarized below.
- The resonance frequency of the mirror-mount-pedestal assembly is 335 Hz (and possibly at higher frequencies?). The peak around 130 Hz is not from the assembly.
- The resonance frequency of the optical table is approximately 119.5 Hz. However, the peak usually observed at 120 Hz has a much narrower bandwidth, suggesting it may not be the resonance of the optical table (possibly power supply noise?).
- The resonance frequencies of laser displacement sensor are 155 Hz and 260 Hz.


Next


・Transfer function measurement using a shaker and accelerometer (Is it necessary?)
・Background measurement with a SUS block as the target
・Cleaning, packing, and dismantling (Setup put by IFI chamber as well)
・Others?
Images attached to this report
VAC (General)
takahiro.yamamoto - 2:59 Friday 10 January 2025 (32290) Print this report
Comment to Slack notification about closing GVs (32289)

I tried to reproduced a change in the pressure value (KGRVAC:CCG_BSY_01:VAL:PRESS) after it became larger than the threshold of guardian notification from the guardian log (See Fig.1).
Log is recorded only when pressure value was larger than the threshold of guardian notification and pressure value is changed, so sampling interval is not a constant value.

Guardian threshold was set as 8.0e-6Pa in klog#32116. The first log appears at 10:55:35.885674 JST. That is, the time when the pressure value first reached the Guardian threshold. Interlock threshold which was shown as red line in Fig.1 was set as 9.4e-6Pa in klog#32173. Readout value first reached interlock threshold at 10:57:52.902407 JST. It took 147s to increase from 8.0e-6Pa (Guardina threshold) to 9.4e-6Pa (interlock threshold). Fig.2 is a same plot with longer duration. Attached text file is an original guardian logs used to make these plots.
 

Images attached to this comment
Non-image files attached to this comment
VAC (Valves & Pumps)
shinji.miyoki - 0:21 Friday 10 January 2025 (32294) Print this report
Comment to GVbsx and GVbsy have been closed (32286)

Owing to two TMP operations near BS, the vacuum level at the MICH area seemed to decrease and reached the equilibrium pressure? This means that more TMPs operation will result in the lower pressure.

The pressure difference at BS before and after closing GVs is just 2 times. 

So can we open GVs? or shall we temporally operate all TMPs in the MICH area and wait to reach the same vacuum level?

Images attached to this comment
VIS (IY)
kenta.tanaka - 22:06 Thursday 09 January 2025 (32292) Print this report
ITMY photosensor balancing was done but ITMY MN sensor diagonalization is on going

To solve the coupling issue in klog, we started to perform the sensor balancing and diagonalization for ITMY sensors.

ITMY photosensor balancing was performed with the script. Fig.1 and Fig.2 show the results in the horizontal and vertical sensors, respectively. They seem to be fine.

After that we performed the ITMY MN sensor diagonalization with the script. Fig.3,4,5,6,7, and 8 show the diagonalization results. They seem to be fine except for the coupling from Y to other DoFs in Fig,8.

I found that the values of the coupling ratios at the Y resonace frequency, 3.11 Hz which were picked up by the script (the last column in the left panel of fig.9) seem to differ from the measument results themselves (the last column in the right panel of fig.9) as for only Y. I tried to input the actual measurement results but the situation was not changed. We need more investigation.

Images attached to this report
VAC (General)
takahiro.yamamoto - 20:10 Thursday 09 January 2025 (32289) Print this report
Slack notification about closing GVs
Because a code implementation became complicated, I had postponed to enable the alert about closing GVs until preparing an alias name as K1:*** instead of KGRVAC:***.
But we faced a trouble related to GVs today (See also klog#32286) and we couldn't catch enough information to improve the interlock system.
So we decided to enable the alert about closing GVs as a mitigation way (NOT a solution) of losing information with the non-DAQ situation.

It works fine for now but current implementation is temporal one for non-K1:*** channels same as klog#32116.
Comments to this report:
takahiro.yamamoto - 2:59 Friday 10 January 2025 (32290) Print this report

I tried to reproduced a change in the pressure value (KGRVAC:CCG_BSY_01:VAL:PRESS) after it became larger than the threshold of guardian notification from the guardian log (See Fig.1).
Log is recorded only when pressure value was larger than the threshold of guardian notification and pressure value is changed, so sampling interval is not a constant value.

Guardian threshold was set as 8.0e-6Pa in klog#32116. The first log appears at 10:55:35.885674 JST. That is, the time when the pressure value first reached the Guardian threshold. Interlock threshold which was shown as red line in Fig.1 was set as 9.4e-6Pa in klog#32173. Readout value first reached interlock threshold at 10:57:52.902407 JST. It took 147s to increase from 8.0e-6Pa (Guardina threshold) to 9.4e-6Pa (interlock threshold). Fig.2 is a same plot with longer duration. Attached text file is an original guardian logs used to make these plots.
 

Images attached to this comment
Non-image files attached to this comment
VAC (Valves & Pumps)
shinji.miyoki - 18:12 Thursday 09 January 2025 (32288) Print this report
Comment to GVbsx and GVbsy have been closed (32286)

Because Kimura-san started two TMPs/Dry pumps near BS(Kimura-san? will report) for the pumping in the MICH area, I,

  • turned off two Sunrise heaters near SR2+X side and PR3-Y side,
  • changed the target temp of the precision cooler near SR2 from 20.00 to 19.00.

to adjust the temperature around BS at around 18:00.

VIS (IX)
takafumi.ushiba - 17:11 Thursday 09 January 2025 (32287) Print this report
Modal damping of ITMX GAS filters

I implemented modal damping for ITMX GAS filters with the same manner as ETMY (klog31916) and ETMX (klog31933).
Figure 1 shows the spectra of LVDTs and OpLevs with (red) and without (blue) modal damping.
Spectra of F3 GAS filters became worse when engaging modal damping but it is not due to modal damping but the ADC glitches, so it doesn't matter.
Other than F3, the spectra became better, so modal damping seems to work well.

I implemented it into te guardian, so M1 modal damping is automatically turned on when the suspension is LOCK_ACQUISITION state.

Images attached to this report
VAC (Valves & Pumps)
takashi.uchiyama - 16:42 Thursday 09 January 2025 (32286) Print this report
GVbsx and GVbsy have been closed
2024/01/09

The interlock system closed GVbsx and GVbsy around 11 AM. As GV closed, the pressure at the center area (IMC- BS- OMC) increased, and the pressure in IXC and IYC decreased.
The reason for the activation of the interlock system is unknown. It is possible that there was a malfunction of the vacuum gauge used for the monitor or a pressure increase due to leaks, etc.

To open GVbsx and GVbsy, we will monitor the pressure change and Qmass measurement until noon tomorrow and judge the results.
Comments to this report:
shinji.miyoki - 18:12 Thursday 09 January 2025 (32288) Print this report

Because Kimura-san started two TMPs/Dry pumps near BS(Kimura-san? will report) for the pumping in the MICH area, I,

  • turned off two Sunrise heaters near SR2+X side and PR3-Y side,
  • changed the target temp of the precision cooler near SR2 from 20.00 to 19.00.

to adjust the temperature around BS at around 18:00.

shinji.miyoki - 0:21 Friday 10 January 2025 (32294) Print this report

Owing to two TMP operations near BS, the vacuum level at the MICH area seemed to decrease and reached the equilibrium pressure? This means that more TMPs operation will result in the lower pressure.

The pressure difference at BS before and after closing GVs is just 2 times. 

So can we open GVs? or shall we temporally operate all TMPs in the MICH area and wait to reach the same vacuum level?

Images attached to this comment
shinji.miyoki - 13:49 Friday 10 January 2025 (32299) Print this report

Because TMPs will be off, I recovered the target temp for the precision cooler from 19 to 20 and turned off two Sunrise heaters around 14:00.

nobuhiro.kimura - 14:12 Friday 10 January 2025 (32300) Print this report

[Kimura and Yasui] 

GVbsx and GVbsy were opened in the following steps.

2Step 1: 

 13:45 Disabled the enable/disable switch in the GVbsx and GVbsy interlock device.

Step 2

13:48 Opened GVbsy and closed the GVs of the TMP.
BS-CC-10 pressure was 1.4 x 10^-5 Pa.

Step 3

13:50 Opened GVbsx and closed the GVs of the TMP.
BS-CC-10 pressure was 1.4 x 10^-5 Pa.

Results of the partial pressure measurement are as follows;

 

BS

IXC

IYC

Time

14:03

14:09

14:05

T.P. (Pa)

1.45 x 10^-5

 4.15 x 10^-6

5.73 x 10^-6

H2 (Pa)

2.00 x 10^-6

 1.68 x 10^-6

2.27 x 10^-6

H2O (Pa)

2.09 x 10^-6

 1.16 x 10^-6

1.62 x 10^-6

N2 (Pa)

5.54 x 10^-7

 5.61 x 10^-7

7.03 x 10^-7

O2 (Pa)

3.54 x 10^-7

 2.07 x 10^-7

2.48 x 10^-7

nobuhiro.kimura - 14:54 Friday 10 January 2025 (32301) Print this report

[kimura and Yasu]

 We turned off two TMPs and two dry-pumps at 14:50.

 

nobuhiro.kimura - 2:47 Saturday 11 January 2025 (32306) Print this report

[Kimura and Yasui]

 BS CC-10 pressure reached 7.5 x 10^-6 (Pa) at 16:10. 

The threshold of the CC-10 contact output was then set to (7.5 + 0.5) x 10^-6 (Pa) of 8.0 x 10^-6 (Pa). 

After setting, the interlock device was enabled.

 

nobuhiro.kimura - 7:56 Saturday 11 January 2025 (32312) Print this report

 

Results of the partial pressure measurement at 7:43 on Jan. 11 are as follows; 

Place

 BS

BS

IXC

IYC

Date

1/07

1/11

1/11

1/11

Time

 10:45

 7:43

 7:39

 7:44

T.P. (Pa)

 1.28 x 10^-5

 1.20 x 10^-6

 4.37 x 10^-6

 4.62 x 10^-6

H2 (Pa)

 1.42 x 10^-5

 1.52 x 10^-6

 1.65 x 10^-6

 2.01 x 10^-6

H2O (Pa)

 1.57 x 10^-6

 1.49 x 10^-6

 1.27 x 10^-6

 1.22 x 10^-6

N2 (Pa)

 6.18 x 10^-7

 5.05 x 10^-7

 6.22 x 10^-7

 6.04 x 10^-7

O2 (Pa)

 3.04 x 10^-7

 3.10 x 10^-7

 2.21 x 10^-7

 2.03 x 10^-7

 

The partial pressure of the residual gas components were recovered to the level before the interlock device operated and the gate valves closed.

 

DetChar (General)
hirotaka.yuzurihara - 15:28 Thursday 09 January 2025 (32284) Print this report
Automation of the daily monitor for identifying and fitting the line noise

[TSYamamoto, Yuzurihara]

TS Yamamoto-san and I have worked on developing the line monitor for one-day data in which the LSC_LOCK guardian is in the state more than PRFPMI_RF_LOCKED. The related klog is klog#31884.
Today, I implemented the automation of the process. The process of analyzing the one-day data runs on k1det1 and will start every 9:25. In the future, it's better to move the running place to the detchar cluster and manage the jobs by using HTCondor. (It's also the same for the Gauch.)
You can access the latest result from the link or the summary page --> DetChar --> Line Monitor.
The result of 2024/12/30 can be seen in the link. The line fit results can be seen in the link.

VIS (IY)
dan.chen - 5:26 Thursday 09 January 2025 (32283) Print this report
Comment to ITMY is oscillating (32218)

(Work on 1/8)
We found that these oscillations grow over the course of several hours, so we turned on these two control loops (Y3, Y4).
Additionally, we modified the Guardian to enable turning them ON/OFF.

Images attached to this comment
Search Help
×

Warning

×