IOO (IMC)stefan.ballmer - 12:04 Tuesday 26 June 2018 (5382)
Print this reportComment to PMC servo final status - operational now (5372)In klog 5372 I forgot that the TTFSS board has 10dB of gain between the common OUT1 and EXC points. So do not use the previously reported open loop gain data and plots. The new ones are attached:
At the nominal gain setting of +10dB common gain and +10dB of fast gain we have a UGF of 7kHz.
Attached are the correct open loop gain and closed loop gain transfer functions plots, as well as the raw data (frequency, real(OLG), imag (OLG).
Also see klog 5155 for a history of all modifications.