Before starting the installation, we transported the required tools and equipment from the central area to Xend, including a steel ruler, a laser level, slings, and a chain block. We also removed the floor mat in the anteroom of the EXC clean booth and cleaned the floor.
One NCal pylon was moved today. In front of the C-chamber anteroom, the pylon was lifted from its pallet using a tripod and placed on a Bishamon lift table. It was then transported into the anteroom. The protective plastic covering was removed, and both the pylon and the lift table were carefully wiped and cleaned.
Inside the clean area, we moved a workbench that had been located between the anteroom and the installation position. We used some C-clamps, slings, and chain blocks to a beam of the clean booth, and slings and chain blocks were connected to it. The pylon was then moved into the clean booth.
At the installation position, we removed the tape covering the anchor holes and cleaned the area. We also removed the waterproof sheet. We found that there is a boundary or gap in the floor near the installation position, which needs to be considered carefully when adjusting the balance of the pylon.
The pylon was suspended from two directions and moved above the installation position. Shim plates equivalent to those used under the cryocoolers in the cryogenic area were inserted between the floor and the pylon to adjust the inclination.
During this work, we found that the prepared M16 bolts were too long, and therefore the pylon could not be completely fixed to the floor. At present, the pylon is temporarily held by the long bolts and the lifting equipment.
The pylon height and inclination were adjusted using a laser level. The reference height was the mark labeled “Cryogenic center,” which was created on August 22, 2017, and reconfirmed on May 3, 2021. We also confirmed that this reference point is approximately consistent with the center of the arm pipe.
This reference point appears to be the one described in klog16644. According to that klog, ETMX was located 8.5 mm below this reference height. However, we have not yet checked whether this relation was modified in later work. For the NCal installation, the important quantity is not simply matching the ETMX height, but accurately determining the relative position between the NCal and ETMX.
A steel ruler was placed on the 640-mm-diameter intermediate positioning plate on top of the pylon. The laser height was measured at four positions corresponding to the (-y), (+x), (+y), and (-x) sides relative to the pylon center.
The final ruler readings today were:
(-y, +x, +y, -x) = (127.5, 127.5, 126.0, 126.0) mm.
Assuming that ETMX is located at the reference height, the nominal design value is approximately 125 mm. However, future fine height adjustment is planned to be performed using shim plates underneath the NCal unit itself rather than underneath the pylon. Therefore, we decided to keep the current readings slightly larger than 125 mm.
The work was not completed today and will continue tomorrow.
Pictures will appear here: link