DetChar (General)
takaaki.yokozawa - 9:46 Monday 27 July 2026 (37270) Print this report
Check the line noises during O4c
Line noise investigation team found several comb line noises during O4c. See JGWDoc17437.
I checked the LSC related PDs
during O4c full lock (2025/07/08 02:39:26 UTC, red)
No IMC output (2025/08/20 00:00:00 UTC, blue)

The excess blue spectrum than red spectrum in LSC-REFL_PDA1_RF45_{I,Q}_ERR_DQ would come from the setting of whitening filters
red : -12 dB, 2 zero-pole fileters
blue : no whitening filters

Anyway, we cansee several 1 Hz comb lines at the
LSC-REFL_PDA1_RF45_Q_ERR_DQ
LSC-REFL_PDA3_RF45_{I,Q}_ERR_DQ
have some 1 Hz comb lines
And no lines at the POP PDs and 3f PDs.

I will check current situation with changing the whitening filters
Images attached to this report
Comments to this report:
takaaki.yokozawa - 5:49 Wednesday 29 July 2026 (37284) Print this report
I missed the attached file.
Images attached to this comment
takaaki.yokozawa - 5:54 Wednesday 29 July 2026 (37285) Print this report
I also compared the spectrum during the O4c and current (IR shutter close, GR shutters open)
The 1 Hz comb peaks disappeared in current sensitivity curve.

My assumption would be related the clean up campaign for GR/IR coupling, klog36533 .
Images attached to this comment