IOO (IMC)
takafumi.ushiba - 12:03 Thursday 04 June 2026 (37025) Print this report
Add HPBD temperature check decorator in IMC guardian

Recently, high power IMC output with misaligning PRM is often used and sometimes IMC is kept locking with high power over night, which results in making HPBD hot,
So, I discussed with uchiyama-san and decided to add the safety function in the IMC guardian.

If hitting 10W beam for 8 and 24 hours, HPBD temperature becomes 350K and 370K, respectively.
So, I set the threshold at 365K as shown in fig1.

From now, if the HPBD temperature exceeds 365K, IMC lock will be lost once to make IMC output smaller.

Note:

During the work, I noticed we have two ZERO_SERVOOFFSET state with the same index.
Since this state is not used, I didn't change them but it would be better to renew the IMC guardian.

Images attached to this report
Comments to this report:
tomotada.akutsu - 21:12 Thursday 04 June 2026 (37030) Print this report

By the way, there are several thermometers attached around the high power beam dump. One of them would be on the top surface of the optical table in the IFI chamber, probably. How about their values?

takafumi.ushiba - 9:26 Friday 05 June 2026 (37034) Print this report

Optical table keeps the temperature around 297K even when HPBD temperature reaches around 365K.