PEM (Center)takaaki.yokozawa - 6:24 Tuesday 17 March 2026 (36594)
Print this reportTapping test at PSL roomI performed the tapping test again. Analysis results would be appear soon.
Comments to this report:
takaaki.yokozawa - 12:19 Tuesday 17 March 2026 (36598)
Print this reportVery preliminary results. Only for the mirrors before the PMC
Fig.1. overview (temporary, I tagged M01' - M04' for the mirrors of the new laser) Fig.2. - Fig.5. showed the various DQ channels. including the PMC related channel, IMMT1 trans QPD2, IMC refl QPD1,2 (DC, 14I, 14Q), IP QPD1,2 and accelerometers
As you can see, Quite large excess can be seen in IMMT1 trans QPD, when we tapped the M3' and M4' mirrors, that the interesting issue.
Images attached to this comment
takaaki.yokozawa - 13:36 Tuesday 17 March 2026 (36600)
Print this reportI just made the all plots for the tapping tests JGWDoc17238