VAC (Valves & Pumps)
nobuhiro.kimura - 20:15 Wednesday 11 March 2026 (36549) Print this report
Switching from TMP to Ion Pumps

[Kimura and Yasui] 

 On the afternoon of March 11, the vacuum evacuation pumps for MCE, #7 (between IXA and IXC), and #8 (between IYA and IYC) were switched to Ion pumps, and TMP and Dry were stopped.
This may cause a change in temperature within the central mirror room.

Tomorrow, we plan to switch the vacuum exhaust pumps on the 1F of the X-end and Y-end from TMP to Ion pumps.

Comments to this report:
nobuhiro.kimura - 5:40 Friday 13 March 2026 (36566) Print this report

[Kimura and Yasui]

On March 12, we switched the vacuum pumps in the central mirror chamber to ion pumps.
The TMP and dry pumps currently in operation are pumps #40 and #41 directly below the arm-side GV.
All other TMP and dry pumps in the central mirror room, including those on the 2nd floor, have been stopped.
The vacuum pumps at X-end 1F have also been switched to ion pumps.
Locations where the switch to ion pumps is not yet complete are the EXVs at X-end 2F and the EXVs at Y-end 1F and 2F.
The switch is scheduled for early next week.

Images attached to this comment
takashi.uchiyama - 13:40 Friday 13 March 2026 (36571) Print this report
2026/03/13

SawadaH, Kimura, Uchiyama

Weekly vacuum system inspection found that TMP(#40) was stopped unexpectedly. We found the interlock system for TMP(#40) was turned off and this was the reason of why TMP(#40) was stopped.

Yesterday, we stopped TMP(#8) and the power breaker for TMP(#8) was turned off. Since the power cable of the interlock system for TMP(#40) was connected to the power breaker for TMP(#8) wrongly, the interlock system was also turned off.

We re-connected the power cable of the interlock system(#40) to the power breaker for TMP(#40) and then restarted TMP(#40).
Images attached to this comment