CAL group
We did the calibration measurements and haven't found strange results.
Estimated parameters in the Pre-maintenance measurements are as follows.
H_etmxtm = 3.87175e-14 @10Hz ( -0.1% from previous measurements)
H_etmxim = 1.55407e-14 @10Hz ( -5.5% from previous measurements)
Optical_gain = 2.187716e12 ( -0.5% from previous measurements)
Cavity_pole = 17.99656 Hz ( -0.4% from previous measurements)
Previous values are listed in klog#34048.
Fig1. shows the fitting results of pre measurements.
Estimated parameters in the Post-maintenance measurements are as follows.
H_etmxtm = 3.878432e-14 @10Hz ( 0.17% from pre-maintenance measurements)
H_etmxim = 1.589672e-14 @10Hz ( 2.3% from pre-maintenance measurements)
Optical_gain = 2.191044e12 ( 0.15% from pre-maintenance measurements)
Cavity_pole = 18.04787 Hz ( 0.28% from pre-maintenance measurements)
Fig2. shows the fitting results of post measurements.