The temperature of EX_50K_REFBRT_HEAD is still fluctuating around 110 K, as shown in Fig.1
On the other hand, the vacuum level at EXT was almost recovered at 2x10^-5 Pa from 10*10^-5 Pa, even though all the BRT1/2/3/4 temperatures were over the critical temperature (Fig.2). The vacuum level enhancement was dominated by water according to the Qmass. In addition, the pressure enhancement was higher at this time than in the last accident, maybe because of the higher temperatures at the BRT1/2/3/4 positions.
The TM_Oplev_LEN_SUM_OUTPUT shows strange changes at the last, and this time accident. The 2/3 reduction in the last accident, but almost a recovery of value in this time. Any frosting on the windows on the inner/outer radiation shields, LEDs, and QPD photosensors in the cryopayload cannot explain the recovery phenomena. On the other hand, the MN_Oplev_LEN_SUM_OUTPUT seems not to show drastic changes at two accident times.
The GRX_NORM seemed to be roughly constant in the last 50 days, while the GRY_NORM showed 1.1! in the past, but around 1.0 at present. Anyway, not so serious frosting on HR sides are expected from this data.