PEM (Center)takaaki.yokozawa - 10:25 Thursday 03 April 2025 (33207)
Print this reportFirst trial of the magnetic injection using large coil[YokozaWashimi]
After this work (klog32939), we performed the first magnetic injection using the large coil. But, we cannot find any coherent signal from DARM (10, 20, 40, 50, 100, 500 Hz) Fig.1. showed the example of the 10 Hz.
Images attached to this report
Comments to this report:
tatsuki.washimi - 16:49 Monday 07 April 2025 (33280)
Print this report
I analyzed the magnetic injection data on April 3rd, based on the coupling function model.
Fig.1: ASDs of the strain and the magnetometer (at BS), for the reference and each injection time