The caution message on chiller for laser

Today, when I walked behind the PSL room, I found the water chiller showed the message `FLtrS`. In the case that this is out of nominal state, I'm writing this in klog. It might be better to check the message by the person in charge.

Images attached to this report
Comments to this report:
osamu.miyakawa - 16:25 Friday 17 January 2025 (32379) Print this report

Miyoki-san noticed the water level was low, so I added 2L of water this afternoon. Now water level is closer to MAX LEVEL.

I did not notice the meanning of the message that Yuzurihara-kun pointed out. I will check it next time when I enter the mine.

osamu.miyakawa - 12:01 Friday 24 January 2025 (32450) Print this report

I checked the status of the water/air filters. The water filter was OK, but the air filter had some dust. I cleaned the surface of the air filter. The "FLtrS" message just counts the running time of filters and I reset the running time to zero. However, it was not possible to disable this "FLtrS" message even though I followed the manual. Probably it requires shutdown once or resetting all the setup. I just leave it this time.