Reports 1-1 of 1 Clear search Modify search
LAS (bKAGRA laser)
Taishi Sako - 17:57 Tuesday 24 January 2023 (23685) Print this report
Converter installing and PZT resonant checking.
1, I installed modified Dsub-BNC converter into ISS lack.
the converter has twice gain than normal one.
As a result, the PZT adjustment range has been expanded.
2, I have input white noise to check the resonance of the PZT.
I want to increase the resonance frequency and reduce the peak because PZT resonance can interfere with control.
I was able to increase the 395Hz resonance a little by tightening the screws tighter.
I was also able to lower the peak by inserting rubber under the mirror mount, but it became unstable, so I removed it immediately.
Images attached to this report
Comments to this report:
Taishi Sako - 18:08 Tuesday 24 January 2023 (23687) Print this report
removed normal converter(S2214289) was return to mozumi office.
modefied converter number is S2214278.