IOO (OMC)Shingo Fujii - 16:43 Tuesday 04 October 2022 (22328)
Print this reportMeasurement of the whitening filter's input voltage and output counts Date: 2022/10/4
With Dan Chen
We measured input voltage vs output counts of the whitening filter in channel 1 and 2.
Images attached to this report
Comments to this report:
dan.chen - 6:40 Wednesday 05 October 2022 (22344)
Print this report
Because these results are consistent with our expected conversion factor 2^16/40V = 1638.4 [count/V], we believe there is no unexpected behavir which can make the signal half reported on klog22313.