Reports 1-1 of 1 Clear search Modify search
IOO (OMC)
Shingo Fujii - 16:43 Tuesday 04 October 2022 (22328) Print this report
Measurement of the whitening filter's input voltage and output counts
Date: 2022/10/4

With Dan Chen

We measured input voltage vs output counts of the whitening filter in channel 1 and 2.


Images attached to this report
Comments to this report:
dan.chen - 6:40 Wednesday 05 October 2022 (22344) Print this report

Because these results are consistent with our expected conversion factor 2^16/40V = 1638.4 [count/V], we believe there is no unexpected behavir which can make the signal half reported on klog22313.

Search Help
×

Warning

×