Lucia, Yoshinori with help of VIS folks and commissioners and ..,
= Background & Target =
--From ITM-height measurement, we found that the height of the IMs were different, and we decided to compensate the difference by the stepping motors (fishing-rods, FR) on the GAS filters.
-- According to the height measurement, we concluded that ITMX should go down by 5.2mm and ITMY should go up by 6mm.
-- Then after the GAS hight adjustment, the height of the TM will be checked interferometer-wise.
= What we did =
-- We updated the calibration factors since some of the calibration factors looked not-updated (ref: klog#8862). sorry for this.
-- We adjusted the GAS-keystone height with FRs;
-- We first checked the available range of each keystone height according to the calibration curves, and then did the adjustment work.
== ITMY, done ==
-- We selected F2 and F3, and pulled up their keystones by ~5.9mm in total (based on the vertical sensor on the BF-damper, K1:VIS-ITMY_BF_DAMP_V_INMON). (Fig1)
-- Since the calibration factor of F2-GAS was changed around +1650counts (about 1V, as shown in klog#5494) and it was not compensated in this work, it looked that the BF-damper displacement was inconsistent with the shift of F2-GAS.
-- The adjusted amout is:
channel name | start position | end position | reference |
K1:VIS-ITMY_F2_DAMP_GAS_INMON (in um) | -1406 | 73 |
K1:VIS-ITMY_BF_DAMP_V_INMON: 324 -> 5000 (del = 4676 um) --(1) |
K1:VIS-ITMY_F3_DAMP_GAS_INMON (in um) | -2200 | -793 |
K1:VIS-ITMY_BF_DAMP_V_INMON: 5000 -> 6272 (del = 1272 um) --(2) |
* (1)+(2)~5.9 mm.
== ITMX, NOT yet done ==
-- We selected F2, F3 and BF, and tried to pull down their keystones by 5.2mm in total, but could pulled down by only ~1.6 mm today. (Fig2)
-- we failed to actuate F3-FR, hopefully due to the driver. This have to be checked in the tunnel.
-- BF-GAS-keystone hit the stopprt frame when K1:VIS-ITMX_BF_DAMP_GAS_INMON was equal to -2000 (um).
-- The adjusted amout is:
channel name | start position | end position | reference |
K1:VIS-ITMY_F2_DAMP_GAS_INMON (in um) | -1297 | -1807 |
K1:VIS-ITMX_BF_DAMP_V_INMON: 2588 -> 1893 (del = -695 um) --(3) |
K1:VIS-ITMY_BF_DAMP_GAS_INMON (in um) | -895 | -1813 | X, del ~ -918 --(4) |
* (3)+(4)~ -1.6 mm.
=== ITMX, Next ===
-- check the status of the stepping motor driver and activate F3-FR.
-- summarize the factors, issues we had so far, and tidy up the information.