mark.barton - 11:18 Friday 03 May 2019 (8799) Print this report
SRM Venting Report

Wednesday 5/1 and Thursday 5/2.

The central area was vented in two phases, on Wed afternoon and all day on Thu. I monitored the SRM (and other Type Bs - see separate reports).

Before:

  • Guardian IM_DAMP_ENGAGED

  • IP SET -400/-50/-3600

  • IP BLEND same

  • IP COILOUTF -3790/-495/3248 (OK)

  • GAS SET -1600/0/0

  • GAS LVDTINF same

  • GAS COILOUTF -6670/-4733/-2400 (OK)

  • OLSET -10/-174 (but OLDCTRL off; OPTICALIGN also off)

  • OLSUM was 17400/13327 (good)

  • OLDIAG was 283/-10/-174. 

  • Initial IF FR 360648/495352/?? (I wrote down the F0 stepper value by mistake instead of #3)

During venting I monitored the IP DCCTRL and COILOUTF values and applied 5000 steps of L or T, or 2000 steps of Y at a time to keep the COILOUTF values less than 5000 cts.

The final IP FR positions were 502908/447092/495000. This was a fairly large change compared to BS.

The F0 COILOUTF went over -20000, so I reduced the SET value from -1600 to -1500 µm. This should be undone after pumpdown. Also, the 100 µm adjustment made a suspiciously large change to the control output. We should check for touching. The  final GAS COILOUTFs were -12400/-11769/-6756.

Comments to this report:
mark.barton - 13:42 Friday 03 May 2019 (8803) Print this report

See 8804.

mark.barton - 14:01 Friday 03 May 2019 (8804) Print this report

The suspension had been fine when I went home after venting finished on Thursday, but Enomoto-san reported that he found the optic had been oscilating in pitch, apparently due to the fact that the TM coil driver WD had tripped. He reset the CD at about 1:45 UTC (10:45 JST) and reset the Guardian to IM_DAMP_ENGAGED. This would have been before I tweaked the F0 GAS SET value and wrote the klog.