[Enomoto, Sugimoto]
One more probrem about Y Prometheus.
We performed a temperature control test, which use BNC port behind the laser controller, but it did not work.
when the knob on the front of the controller was operated directly, normal operation was confirmed.
The operation confirmation was carried out by measuring the frequency fluctuation of the beatnote signal by the main laser and Prometheus.
What we tried (didn't work)
-
Tried a temperature control operation from MEDM by connecting DAC of ALS rack and laser controller PSL .
-
Confirmed that the voltage output from the DAC was normal, and checked the voltage even at the BNC on the PSL side.
-
A sinusoidal wave of 0.1 Hz,50 mV was added to the laser controller by a function generator.
I performed a temperature control test, which use BNC port behind the laser controller, but it did not work.
when the knob on the front of the controller was operated directly, normal operation was confirmed.
The operation confirmation was carried out by measuring the frequency fluctuation of the beatnote signal by the main laser and Prometheus.
What we tried (didn't work)
.Tried a temperature control operation from MEDM by connecting DAC of ALS rack and laser controller PSL .
.Confirmed that the voltage output from the DAC was normal, and checked the voltage even at the BNC on the PSL side.
.A sinusoidal wave of 0.1 Hz,50 mV was added to the laser controller by a function generator.