Michimura, K. Tanaka, Tanioka
Today, Michimura-san noticed that the configurations of QPD segments and seising matrices diaplayed on MEDM screen were incorrect.
So we investigated their configuration by using a laser pointer and modified.
The segments configurations of MCE TRANS QPD and IMMT1 TRANS QPD1&2 shown on the medm were wrong.
Also, the sensing matrices of IMMT1 TRANS QPD 1&2 were not correct.
Especially, the plus-minus sign of PIT was wrong.
We modified them (medm screen and sensing matrices).
The correct version is shown in the 2nd figure.