I am so sorry that I have to report this accident.
Today I accidentally broke RFPD (JGW-D1201280) for PMC REFL. The reason why it was broken is presumably that -18 V voltage supply was turned off (my foot touched the switch lever of the supply but I didn't notice that at that time) and was left off for a few minutes, even though +18 V was on. Then Nakano-san and I found that both DC and RF output signal from the RFPD behaved strangely.
Awai-san and I surveyed the circuit of RFPD and finally found that at least two Op. Amps. in the DC output line were broken.
For the time being, an RFPD from CLIO will be used for acquisition of the error signal of PMC.
Supplying only +18V (or -18V) would damage RFPDs seriously, so please be careful of DC voltage supplies.
I hope no one will cause the same kind of accident.