Reports 1-1 of 1 Clear search Modify search
IOO (IFI)
keiko.kokeyama - 19:21 Wednesday 29 August 2018 (5943) Print this report
bad clamp hunting

[Enomoto, Kokeyama]

Some of the clamps are used badly for the IFI. Dog clamps must push the fixing point from higher height than the pushing point.

We removed those clamps and replaced. The picture shows a part of what we done.

Changing list:
- IFI output baffle clamps x 2
- IFI input baffle clamp 1 (one of the two cannot be swapped because of the limited space between the base place and the baffle leg)
- IFI middle baffle clamps x 2
- Glass beam dump (same clamp, but adding a washer for more height of the supporting point)
- STM1 clamps x 2 (washers added for more height)
- POM clamps x 2 (washers added for more height)

To do list:
- One more clamp with a washer for STM1
- Two clamps for STM2 (with two washers)
- One more clamp with washer for the glass beam dump

BEFORE


AFTER
   

Used clamps are commertial, Sigma Koki Vacuum compatible ones.
 

Images attached to this report
Search Help
×

Warning

×