VIS (BS)
kiwamu.izumi - 22:17 Monday 23 April 2018 (4799) Print this report
IM and TM responses re-measured

Related logs: 4793, 4776

We re-measured the TM and IM responses of the BS tonight with the MICH loop closed.

They are now calibrated in to [m/cnts] using the estimated optical gain of 5e-10 m/cnts (4795). Comparing the data against those of the day before yesterday (4776), the IM response was found to be greater today as a consequence of disabling the low-pass filter in the coil driver. In other words, the measurement the other day contained the unintentional low pass consisting of a pole at 1 Hz and zero at 10 Hz. The measurement today should be with the flat response in the coil driver instead.

The second attachment is the calibrated actuator responses in ASCII format.

Images attached to this report
Non-image files attached to this report