Reports 1-1 of 1 Clear search Modify search
FCL (Air)
shinji.miyoki - 10:26 Tuesday 17 November 2026 (37848) Print this report
Tehrmometer Change and #1 Precision cooler operation start

[miyoki, miyakawa]

Abstract

#1 precision cooler operation started. The position of its reference thermometer was changed. The PID parameters of the #1 precision cooler were adjusted. Then we stopped #2 precision cooler for the near-future inverter maintenance.

 

Background

Error 093 was shown on the #2 precision cooler. This means the inverter's remaining life. Before changing the inverter, we decided to operate the #1 precision cooler, which had been off for a long time, instead of the #2 precision cooler, which had been in use for a long time.

 

Process

For laser safety, the laser was turned off by Miyakawa-kun during my work. Off the amp, off the master laser. (We did not remember whether the noise eater was on or off.)

To operate #2 and to stop #1, we need to change the thermometer position that is used for the temperature reference to keep the constant temperature injection, from at the side of the PMC to at the duct outside the PSL room as #1 (klog#28317). Also, we need to adjust the PID parameters of #1. According to the PID settings of #2, PID = 9.9, 72, and 18. So we changed them for #1 from 7, 45, and 10 to 9.9, 72, and 18., temporally.

We replaced the thermometer position on the air injection duct from the #1 precision cooler as in photos 1, 2, and 3.

We restarted the #1 precision cooler initially with 21.09C target temperature that is the same to the #1's. Before that, we opened the entrance of the PSL room to avoid too much pressure inside the PSL room by operating both #1 and #2 precision coolers.  After the temp stabilization of the air from the #1 precision cooler, we stopped the #2, closed the PLS room's door again, and waited for the stabilization of temps on the PSL optical table. Because the stabilization performance seems to have oscillations, I performed auto PID parameter tuning. Then the parameters became 1.8, 71, and 18.

We are still monitoring the PT01~ PT06 temperatures.

Additional Note

Fujimoto-kun and Hirose-san reported that the PMC trans power decreased a lot! We noticed that the PT04 thermometer, which is set beside PMC, showed a decreasing trend from around AM10:00 by 0.1C. Our work started around PM1:30. According to YamaT-san, he suspected that the PMC body's temperature decreased without heating up in the PMC control. So, one possibility is deformation of the PMC.

Images attached to this report
Comments to this report:
shinji.miyoki - 10:26 Tuesday 17 November 2026 (37858) Print this report

When we scan the PMC length by PZT, two or three peaks can be seen when TEM00 resonates. Actually, all higher-mode resonances seem to have double (triple) peaks. 

So, some strong frequency modulation might be added from somewhere. 

shinji.miyoki - 10:27 Tuesday 17 November 2026 (37861) Print this report

The figure showed the temp changes of PT01 ~ 06. 

  • All temp started decreasing after the start of DGS maintenance.
  • PT04, which was touched with the replaced thermometer, showed an instant increase because of my finger touch.
  • All PTs showed increasing by 1) DGS recovery and the target temp of #1 from 21.1 to 21.5 to speed up the temp recovery.
  • I reduced the target temp from 22.5 to 22.2 to compensate for too much heat-up. 
Images attached to this comment