ISC (General)
takafumi.ushiba - 17:14 Friday 13 November 2026 (37840) Print this report
Comparison of RFPD signals with MICH excitation before and after MC length tuning

Abstract:

MICH signals on REFL 51Q seems reducing by a factor of 10 compared to the values before MC length tuning.

Detail:

I compared the ratio between POP 17Q and REFL51Q/REFL135Q while exciting MICH DoFs (fig1).
Following table shows the signals (preak height at excitation frequency) of PDs during the measurement.

  November 13 (after MC length tuning) August 16 (before MC length tuning)
POP17Q 0.028952 0.0127529
REFL51Q 0.00306384 0.0133902
REFL135Q 0.0165872 0.00522424

From these results, I calculated the ratio between POP 17Q (we thought MICH signals are dominant on this signal) and the others.
The result is as follows:

  November 13 (after MC length tuning) August 16 (before MC length tuning)
REFL51Q/POP17Q 0.1058 1.050
REFL135Q/POP17Q 0.5729 0.4097

So, the ratio between REFL51Q and POP17Q was reduced by a factor of 10 after MC length adjustment.
On the other hands, ratio between REFL135Q and POP17Q was slightly improved.

We are not so sure why this signal reduction happens, so further investigation is necessary.

Images attached to this report