VAC (EX)
nobuhiro.kimura - 8:26 Wednesday 04 November 2026 (37796) Print this report
Vacuum leak test for EXC on Nov. 2

[Kimura and Yasui]

 On Nov. 2, we conducted a vacuum leak test on the EXC.
The results showed that no vacuum leaks were detected at a sensitivity of 1×10^-10 Pa·m³/s at the re-tightened flanges of the EXC.
However, since doubts arose regarding the detection sensitivity of the leak detector used, we decided to conduct a functional test on the leak detector.
Depending on the results of the functional test, we may conduct the EXC vacuum leak test again.

Comments to this report:
nobuhiro.kimura - 21:22 Monday 16 November 2026 (37855) Print this report

[Kimura and Yasui]

 On November 16, we conducted another vacuum leak test on the EXC. For this retest, we used a helium leak detector, which differed from the one used in the previous test.
As a result, no vacuum leaks were detected at the flange section of the EXC that had been retightened, at a sensitivity of 1×10^-10 Pa·m³/s.
 Based on the above, we determined that the vacuum flange of the retightened EXC is leak-free and in good condition.