Kohei Mitsuhashi - 19:19 Friday 02 October 2026 (37604) Print this report
Preparation for the ETMX triangulation survey
[Mitsuhashi Dan]

In preparation for the triangulation survey, we brought the total station to the site and carried out setup and assembly work.

A detailed work log will be added later.
Images attached to this report
Comments to this report:
Kohei Mitsuhashi - 23:12 Saturday 03 October 2026 (37607) Print this report

Preparation for the ETMX Position Survey



Background and purpose



In the NCal calibration, the uncertainty in the center-to-center distance between the NCal and the test mass is one of the major contributors to the overall uncertainty of the NCal signal.

Therefore, while the ETMX chamber is open, it is important to directly survey reference points near the test mass and thereby reduce the uncertainty in the test-mass center position compared with an estimate based only on design drawings and manufacturing tolerances.

Prior to the actual ETMX survey, the surveying instruments were brought to the X-end, and the instrument setup and basic surveying procedure were tested. In particular, an object with a known length was measured to verify the procedure for reconstructing the distance between two points from the measured horizontal angles, vertical angles, and slope distances.

Survey preparation



A Leica T30 total station was brought to the X-end and assembled for the survey.

The horizontal-angle convention of the T30 was confirmed to be Horizontal Right (HR), meaning that the horizontal angle increases clockwise when viewed from above.

A KEK sticker attached to the wall was also sighted and measured to confirm that targets on the wall can be observed with the total station under the conditions expected for the actual survey.

As a basic check of the surveying procedure, both ends of a 1500 mm reference scale were measured. The horizontal angle, vertical angle, and slope distance to each endpoint were recorded, and the three-dimensional distance between the two endpoints was calculated from these measurements.

Check of the electronic level



The T30 has a built-in two-axis electronic level, with the two axes displayed as the L and T axes.

Initially, after leveling the instrument such that the electronic-level readings were close to 0°, the total station was rotated horizontally by 180°. A change in the electronic-level readings was observed after the rotation, with a maximum difference of approximately 0.1°.

This indicated that the electronic level had a zero-point offset.

Based on the 180° reversal measurement, the offsets were estimated to correspond to reference readings of 0.0008° for the L axis and 0.0091° for the T axis. The instrument was then leveled more carefully while taking these offsets into account.

After this adjustment, the total station was rotated again over a wide range of horizontal angles. The electronic-level readings remained nearly constant, with variations suppressed to approximately 0.001°. This confirmed that the instrument remained level as it was rotated.

As an independent check, a laser line level was also used to establish a horizontal reference. The telescope of the total station was rotated over a wide range of horizontal angles, and the projected laser line was observed at several positions located at approximately the same distance from the laser level. At these positions, the laser line was found to coincide closely with the center of the telescope field of view.

The agreement between the laser reference and the telescope line of sight over a wide horizontal-angle range provides an additional confirmation that the total station was properly leveled.

Measurement of the 1500 mm reference scale



The horizontal angle, vertical angle, and slope distance were measured for both ends of the 1500 mm reference scale. The distance between the two endpoints was then reconstructed from these measurements.

The calculated distance was 1502 mm, corresponding to a difference of 2 mm from the nominal length of 1500 mm.

Since this measurement was intended as a preliminary check of the instrument setup and surveying procedure, this result alone does not demonstrate a surveying accuracy of 2 mm. However, it confirms that the overall procedure for reconstructing the three-dimensional distance between two measured points from the angular and distance measurements is functioning as expected.
Images attached to this comment
Kohei Mitsuhashi - 21:06 Monday 05 October 2026 (37612) Print this report
[Mitsuhashi, Dan, Araki, Tomaru]

ETMX Position Survey — Day 1

Detailed measurement data and survey results will be added tomorrow or later.

Today, 5 October, we carried out the first day of the ETMX position survey using a Leica T30 total station, following the preparation described in [klog 37607](https://klog.icrr.u-tokyo.ac.jp/osl/?r=37607).

We plan to take measurements from at least two locations. Today, we completed the measurements from one location.

The following target points were observed:

- Four corners on the near side of the sapphire nail heads, as viewed through the chamber opening.
- Four edge points on the copper clamps securing the nail heads.
- Two reference markers attached to pillars outside the chamber, within the clean booth at the X-end.

We also measured the marker for another measurement location to establish a baseline for the triangulation survey.
Kohei Mitsuhashi - 22:53 Tuesday 06 October 2026 (37621) Print this report
[Mitsuhashi, Dan, Araki, Tomaru]

ETMX Triangulation Survey — Day 2



A detailed report on the triangulation survey will be added later.

Following the preparation described in [klog 37607](https://klog.icrr.u-tokyo.ac.jp/osl/?r=37607), we carried out a triangulation survey at the X-end using a total station placed at three locations: A, B, and C.

The following points were observed from these three locations for triangulation:

- Reference markers P1 and P2.
- Four corners of the copper housing, circled in red and named C1–C4.
- Four corners of the sapphire nail heads, circled in green and named S1–S4.

In addition, the following points were measured from location C only:

- The HR-side corner of the test-mass ear.
- The HR-side upper corner of the AR-side sapphire nail head bonded to the ear.

Supplements


We aligned the horizontal line of a green laser level with the top surface of the sapphire nail head and made a best-effort adjustment to align it with the center of the P1 marker.
Images attached to this comment
Kohei Mitsuhashi - 15:18 Friday 09 October 2026 (37638) Print this report

Taking picture ETMX with a scaler using TCam held on 10/6



We taking picture ETMX with a scaler using TCam.
The scaler was put on the duct on AR side, and was on the bottom of the duct.
Tape was attached by 5cm.
Images attached to this comment
dan.chen - 15:45 Friday 09 October 2026 (37640) Print this report

This is a follow-up to the previous klog post.

Background / Purpose

The purpose of this test was to investigate the field of view of the TCam, particularly how much of the area behind ETMX (TMS side) is visible in the TCam images.

By understanding the field of view, we may be able to identify suitable locations for reference markers that could be used to determine the viewing direction of the TCam in the future.

Additional Information on the Ruler

The attached photo shows the 30 cm ruler used in the test.

The white object near the center of the TCam image is the ruler, which was placed at the bottom of the duct on the AR side of ETMX.

Since the ruler's scale markings were not distinguishable in the TCam image, tape markers were attached at 5 cm intervals from the bottom of the ruler.

The upper edge of each tape marker indicates the corresponding 5 cm position.

Images attached to this comment
Kohei Mitsuhashi - 22:38 Saturday 10 October 2026 (37644) Print this report

ETMX triangulation survey: baseline, reference frame and S-point comparison on horizontal plane


We analysed AB face 1/2 and AC face 1/2, comparing the station baselines, P1/P2 reference frame and S1–S4 coordinates.


1. Baseline check — Fig. 1


Using AB = 1268.25 mm and the station directions, the calculated AC length exceeded the measured AC = 2491.5 mm by 0.856 mm in face 1 and 0.761 mm in face 2. The station triangle therefore showed sub-millimetre internal agreement.


2. P1 position and P1→P2 reference direction — Fig. 2


Before alignment, P1 positions and P1→P2 directions were compared in the A-origin/A→B frame. For the three retained reconstructions:


SeriesP1 position difference from AB face 1 [mm]P1–P2 length [mm]Direction difference from AB face 1 [deg]
AB face 10 (reference)3795.8770 (reference)
AB face 22.6413797.467−0.037630
AC face 11.1993794.271−0.004597

The maximum difference in P1–P2 length was 3.196 mm. These are differences between reconstructions; AB face 1 is a comparison reference, not an established true position or direction.


Defining P1 as the origin and P1→P2 as the X axis makes the origin and axis coincide by construction. It does not remove their measurement uncertainty. The MC calculation recomputes P1 and the axis in each trial and propagates their effects into the relative S coordinates.


3. S-point comparison — Fig. 3, right panel


Each reconstruction was transformed to the P1-origin/P1→P2 frame using rotation and translation without scaling. AC face 2 showed the largest overall discrepancies and was excluded from the provisional three-series statistics, while remaining in the diagnostic plots. The responsible observation has not been identified.


PointThree-series sample standard deviation [mm]
S11.93
S21.58
S31.99
S42.06

Values are the largest standard deviations over horizontal distance directions. They describe series-to-series scatter, rather than maximum errors. Because the reconstructions share observations, these values were not divided by √3.


4. Remaining difference and follow-up


Both AC faces show a similar S-row trend that differs from AB. The current MC model does not fully explain the configuration difference, so a common effect in the AC reconstruction remains to be investigated.


S/C targets can no longer be reobserved. We plan to remeasure the accessible ABC markers and P1/P2 in both faces, including the previously missing B→C direction and horizontal baseline checks. Provided the relative control-point geometry is unchanged, the new network will test the old baseline and reference directions and allow identifiable corrections to be applied to the existing S/C data.


The 2 mm survey requirement at 1σ has not yet been demonstrated for all S points.


Images attached to this comment