IR HP laser
- Turned on the chiller for the IR laser → No issues such as water leakage were observed.
- Checked opening/closing the IR shutter in local mode → OK.
- Checked opening/closing the IR shutter in remote mode → OK. After checking, I closed the shutter again.
- Checked the laser power supplies and related equipment → There was no strange behavior.
- Cleared the interlock of laser system from the HP Laser screen.
- Turned on the master laser.
- Turned on the HP laser with 3 W → No abnormalities were observed. Then restored the output to 7.7 W.
PMC
- Set the PMC RF signal generator to 15.25 MHz and 8.6 dBm, and turned on the RF output.
- Turned the HV power supply for the PZT off once, and then turned it on → The voltage returned to around 150 V.
- The PMC Guardian control from the IOO Guardian was set to PAUSE, so it was changed to EXEC.
- Set to "PSL locked” and confirmed that the PMC locked successfully.
The signal generator for the ISS AOM driver was set to 90 MHz and 1 Vpp after the power outage was back. I am not so sure whether they are the correct settings.
In addition, we checked the PZT drivers for the PLL. Ushiba-kun remotely turned the PZT voltage, and I confirmed in the PSL room that the voltage reading increased to around 75 V.