dan.chen - 7:56 Monday 24 August 2026 (37401) Print this report
Initial alignment 260824

With Yokozawa-san

Initial alignment for Xarm, Yarm, PRMI, SRM were performed.

Comments to this report:
takaaki.yokozawa - 3:49 Tuesday 25 August 2026 (37409) Print this report
After the initial alignment, we performed the several trial of DRMI ADS.

- Engaged IMMT2 and PRM ADS
As expected, it was stable

- Engaged IMMT2, PRM and SRM ADS
We tried to engage the IMMT2, PRM and SRM ADS. Before the SRM became stable, lock loss happened.
From the error signal of SRM ADS, the value close to zero, but lock loss happened.
t
- Engaged IMMT2, PRM and BS ADS
Similar with the above situation, BS ADS cannot be stable state, POP 90I increased, but lock loss happened.

- Engaged IMMT2, PRM, BS and SRM ADS

In the situation of this morning, we cannot reach the stable alignment for those ADS, but the direction of the ADS would be fine.

- Engaged IMMT2, PRM, BS, SRM ADS and SRM BPC by SR2
We tired to engage those controls with several gain tweaking, but we failed the stable lock.