thank you for measuring. i have one question. do you compensate the DARM cavity pole for RSE in this measurements? because RSE cavity pole should be changed from PRFPMI one.
Sorry. I'm talking about not ALS_DARM "lock" but the ALS_DARM "sensor". You measured the TFs from EX actuators to ALS_DARM sensor, K1:LSC-ALS_DARM_OUT. This ALS_DARM sensor has a pseudo CP filter to mimic an IR DARM responce. For example, in the PRFPMI case, There is the CP filter, which is labeled as "CP_PRFPMI", in FM2 of this filter bank. In other words, for RSE, you need a filter corresponding to the ALS_DARM_OUT filter bank that is matched to the RSE CP because RSE CP differs from PRFPMI CP. My question is: Have you performed that operation?
But, even if that operation wasn't performed, it's very good to know that the actuators themselves have not changed since PRFPMI.