Reports 1-1 of 1 Clear search Modify search
PEM (Center)
takaaki.yokozawa - 6:24 Tuesday 17 March 2026 (36594) Print this report
Tapping test at PSL room
I performed the tapping test again.
Analysis results would be appear soon.
Comments to this report:
takaaki.yokozawa - 12:19 Tuesday 17 March 2026 (36598) Print this report
Very preliminary results.
Only for the mirrors before the PMC

Fig.1. overview (temporary, I tagged M01' - M04' for the mirrors of the new laser)
Fig.2. - Fig.5. showed the various DQ channels. including the PMC related channel, IMMT1 trans QPD2, IMC refl QPD1,2 (DC, 14I, 14Q), IP QPD1,2 and accelerometers

As you can see, Quite large excess can be seen in IMMT1 trans QPD, when we tapped the M3' and M4' mirrors, that the interesting issue.
Images attached to this comment
takaaki.yokozawa - 13:36 Tuesday 17 March 2026 (36600) Print this report
I just made the all plots for the tapping tests JGWDoc17238
takaaki.yokozawa - 8:45 Wednesday 18 March 2026 (36612) Print this report
Global trend
Similar shape in POW PMC OUT and IP QPD P,Y
large signal in IMMT1 trans if POW PMC OUT became large
Several different in IMC refl QPD 1 and 2
Distance dependence of the response of IP1, 2
1 inch : basically around 500 Hz had resonant frequency
2 inch : basically around 330 and 470 Hz have peaks
150 Hz(M18-1) and 210 Hz (M17) is special resonant frequency

Before the PMC
Relatively small signal in the IMC refl sensors
large peak at POW PMC OUT
340(M04’), 360(M3’) and 680(M04’) Hz may come from the 2 inch mirror before the PMC

After the PMC
M5 1 inch : some 520 Hz signal in IMC refl sensors, several peak in IP pitch signal
M6 1 inch : Quite large peak in 470 Hz in IMC refl sensors, IP peak except for QPD1 yaw ,Small peak in IMMT1 trans
M7 2 inch : some peak in 470 Hz in IMC refl, IP, ACC1, small peak in IMMT1 trans P>Y
M8 1 inch : Peak at 520 Hz in IMC refl and IP, DC QDP2 pitch largest
M9 1 inch : Peak in 510 Hz in various IMC refl (QPD1 larger), and IP (QPD1 larger), 2nd peak 560 Hz in several sensors
M10 1 inch : 490 Hz peaks in several sensors
M18-1 2 inch: large peak at 150 Hz, several peak in 295 and 440 Hz, 640 Hz
M12 2 inch : peak at 330 Hz, 485 Hz(shape different), 550 Hz
M13 1 inch : peak at 550 Hz
M14 1 inch : peak at 510 Hz
M17 2 inch : peak at 210 Hz, 337, 420, 640, 710 Hz
M18-2 2 inch : 337, 350, 550, 680 Hz
M19 2 inch : 365, 395 Hz
M20 2 inch : peaks in 320, 350 and 630 Hz
periscope upper : 450 Hz
periscope lower : 450 Hz