I measured the spectrum ratio (TF Magnitude) of REFL 17/45 QPD1/2 PIT/YAW and POP 17 QPD1/2 PIT/YAW signals with 13.125 Hz Injection on IMMT2/PR3/BS.
- Gurdian State:
INCREASE_SOFTHARD_GAIN - "resG_measure" (FM4) for the LSC DARM2/MICH1/PRCL1 were turned on manually
- Referrence: 2025-10-24 (klog35408)
Following tables are the rough summary of the measured data in dB (original data can be found at /users/Commissioning/data/ASC/2026/0122/).
The values in ( ) are the previous results.
| PIT | REFL_QPDA2_RF17_I | REFL_QPDA1_RF45_I | POP QPDA1_RF17_Q |
| IMMT2 | 50 (48) | 19 (20) | -10 (-53) |
| PR3 | 44 (41) | 45 (61) | 16 (-1) |
| BS | 37 (36) | 30 (46) | 8 (11) |
| YAW | REFL_QPDA2_RF17_I | REFL_QPDA1_RF45_I | POP QPDA1_RF17_Q |
| IMMT2 | 52 (52) | 9 (27) | -6 (24) |
| PR3 | 38 (29) | 47 (59) | 17 (2) |
| BS | 27 (33) | 40 (53) | 14 (16) |