CAL (General)Shingo Hido - 19:12 Thursday 20 November 2025 (35612)
Print this reportCalibration check after O4cCAL group (11/19) We performed calibration measurements to check the current parameters.
Estimated parameters in the previous measurements are as follows. H_etmxtm = 3.908232111e-14 @10Hz ( -0.34% from previous measurements) H_etmxim = 1.487776148e-14 @10Hz ( -2.78% from previous measurements) Optical_gain = 1.891353135e+12 ( -2.48% from previous measurements) Cavity_pole = 18.376366917 Hz ( -0.35% from previous measurements)
No significant deviations were observed compared to the previous measurements. We did not update the parameters.