takahiro.yamamoto - 21:05 Friday 07 November 2025 (35531) Print this report
The Return of the Combs

For checking that the frequency points at the swept sine of calibration measurements are contaminated by spectral peaks, I made a latest ASD with fine (~mHz) resolution.
Then I found that 4.17Hz and 9.77Hz combs came back which were seen also on OMC dark and CARM error signal, respectively as shown in Fig.1 and Fig.2.

These two plots were made with the data around 15:00 JST and 23:00 JST on Nov. 6th. Former and latter locks can be seen in Fig.3 as t = 6h-11h and t = 14h-20.5h, respectively. The lock duration of these two were ~5hr and ~6.5hr and there was no human activities in order to check the stability with disabling the whitening filter stage for the DCPD (probably for former lock and surely for latter lock in my memory).

These two combs were loud during O4a and were mitigated before before O4c (Comparison around 100Hz would be the easiest to understand). I couldn't remember how to mitigate the 4.17Hz combs on OMC (pin assignment of ZSW, cabling route around AS table, or else?). Though there was possibility that noise itself became louder than before, disabling the whitening filter makes SNR of these series peaks louder. If so, there may be no way to suppress them for now because stable IFO is unavailable with engaging whitening filter. In my memory about 9.77Hz combs on CARM, they were able to mitigated by adjusting the input gain of CMS as positive. Though I was not able to track recent activities about replacing IMC-CMS precisely, now the original common mode servo chassis were used for IMC and input gain were set as positive. So I have no idea why 9.77Hz combs came back. replacing chassis and/or re-cabling changed some situation around electrical connections...?

Images attached to this report
Search Help
×

Warning

×