Reports 1-1 of 1 Clear search Modify search
MIF (General)
takaaki.yokozawa - 8:22 Thursday 09 October 2025 (35283) Print this report
Finesse measurement Xarm 251009
After the Xarm alignment, I tried to measure the Finesse Xarm.
But when IFO_REFL_HWP set to 152, locked loss happened when increasing to 3 W.
So I tweaked the IFO_REFL_HWP and target PSL_HWP, I can measure the Xarm finesse

DATE : 2025/10/08 22:44 UTC
ARM : X
TEMP ITM : 90.3
TEMP ETM : 43.4
NUM : 10
IFO REFL HWP : 156.0
PSL HWP : 170.0
IMC POWER : 5.1
VALUE : 1331.3
ERROR : 15.1

Measured data is stored to /users/Commissioning/data/Finesse/Xarm/20251008-224443
Images attached to this report
Search Help
×

Warning

×