Laser frequency noise measurement with PMC

I measured frequency noise spectrum with PMC (fig1).
The measurement was done by locking PMC without IMC and calculated by the feedback signals of PMC PZT.
According to JGW-G2415983, the laser frequency noise was less than 100Hz/rtHz @ 100Hz but measured vaue is about 4 times larger.