CRY (General)
shinji.miyoki - 6:57 Tuesday 15 July 2025 (34559) Print this report
IM Temperature adjustment

Because of high seismic noise due to several big earthquakes, a swarm of earthquakes, and a typhoon (Fig.4), the IFO lock was almost not realized last night. Therefore, the IM temps of ITMY and ETMX decreased significantly. I kept adjusting the current of the heaters for them. The attached values are the original settings before the adjustments. 

Fig.1 and 2 show the original settings for ITMY and ETMX. I increased by 2 Volts for each for several hours. After that, I reduced by 1 volt for each. So the present voltages are larger than 1 volt for each. 

I will recover the original setting values after confirming the stable, continuous lock.

Images attached to this report
Comments to this report:
shinji.miyoki - 15:04 Tuesday 15 July 2025 (34566) Print this report

I recovered the voltage settings for IM heaters for ITMY and ETMX around 14:40 because OBS has started.