takafumi.ushiba - 11:35 Wednesday 09 July 2025 (34495) Print this report
Check GAS output

I checked GAS filter output of all suspensions and found that PR3 SF GAS and SRM F1 GAS is close to the saturation.
Figure 1 shows the GAS filter output at IOP models (left:PR3 SF, right:SRM F1).

In addition, ITMY GAS output is also close to the limit (output is more than 85% of ful range, fig2).
So, it would be better to offload them in the next maintainance day.

Images attached to this report
Comments to this report:
hirotaka.yuzurihara - 11:54 Wednesday 09 July 2025 (34496) Print this report