Reports 1-1 of 1 Clear search Modify search
DetChar (General)
shoichi.oshino - 16:49 Friday 13 June 2025 (34218) Print this report
Hveto 20250612
The glitch around 30Hz and 60Hz had coincidence mainly with ASC-DSOFT_P.
The glitch around 40Hz had coincidence mainly with LSC-AS_PDA1_RF17_Q_ERR.
The glitch from 70 to 90 Hz has many coincidence with LSC-POP_PDA1_RF45_I_ERR.
The glitch from 300 to 400 Hz has many coincidence with IMC-REFL_QPDA1_RF14_I_PIT, ASC-REFL_QPDA1_RF45_I_PIT, OMC-QPDV2_PIT and PEM-ACC_EYA_TABLE_TX_Z.
ASC-AS_QPDA2_DC_SUM_OUT was characterized as having a lot of coincidence at 16 to 22 UTC lock.