Reports 1-1 of 1 Clear search Modify search
MIF (General)
Chia-Jui Chou - 22:12 Tuesday 10 June 2025 (34147) Print this report
Noise Subtraction Results using DeepClean on KAGRA Data

I applied DeepClean to clean KAGRA strain data from 1432093271 s to 1432004295 s (03:40:53, May 24, 2025 UTC ~ 03:57:57, May 24, 2025 UTC, 1024 seconds in total). The DeepClean models are trained using the witness channels and strain channel from 1432094295 s to 1432096343 s (03:57:57, May 24, 2025 UTC ~ 04:32:05, May 24, 2025 UTC, 2048 s in total).

The frequency bands cleaned are: [298Hz, 302Hz], [322Hz, 343Hz], [348Hz, 380Hz], [418Hz, 422Hz], [440Hz, 450Hz], [462Hz, 472Hz], [533Hz, 542Hz]. The witness channels used for each frequency bands can be found here: JGW-T2516710.

The inspiral range of the strain data was calculated using 84 seconds of data starts from 1432093271 s with fftlength=8s, overlap=4s, fmin=10Hz. and we can see the inspiral range got improved by 0.476% after cleaning.

We are working on the cleaning using Independent Compoment Analysis (ICA) method and also trying to improve DeepClean by introducing coherence ratio as part of the loss function.

 

Images attached to this report
Search Help
×

Warning

×