PEM (Center)
takaaki.yokozawa - 9:00 Thursday 10 April 2025 (33337) Print this report
PEM injection test 250410
I performed the PEM injection this morning

Before this measurement, it would be silent run

2025/04/10 6:38:00 - 07:48:00 Acoustic injection PSL room
Excitation : K1:PEM-EXCITATION_MCF0_RACK_13_EXC
50 - 400 Hz , 1Hz resolution, 10 s in each measurement, 100 cnt
Fig.1. showed the whitening spectrogram
Four IPC glitch like signal detected (17, 31, 37, 59 min)
Large signal detected in 65 min, but it may come from another glitch
We need to test the signal around 12 min and 48 min, but they would also come from another glitch, I think.

2025/04/10 8:49:00 - 08:59:00 Acoustic injection PSL room
Excitation : K1:PEM-EXCITATION_MCF0_RACK_13_EXC
300 - 400 Hz , 2Hz resolution, 10 s in each measurement, 100 cnt
Fig.2. showed the whitening spectrogram
Images attached to this report
Comments to this report:
tatsuki.washimi - 16:56 Monday 14 April 2025 (33402) Print this report

I analyzed the injection data of

2025/04/10 6:38:00 - 07:48:00 Acoustic injection PSL room
Excitation : K1:PEM-EXCITATION_MCF0_RACK_13_EXC
50 - 400 Hz , 1Hz resolution, 10 s in each measurement, 100 cnt

 

The IPC glitch times are removed from the analysis, but the other glitches-like data is included.

Images attached to this comment
tatsuki.washimi - 17:41 Monday 14 April 2025 (33406) Print this report

I also analyzed the injection data of

2025/04/10 8:49:00 - 08:59:00 Acoustic injection PSL room
Excitation : K1:PEM-EXCITATION_MCF0_RACK_13_EXC
300 - 400 Hz , 2Hz resolution, 10 s in each measurement, 100 cnt

Here, the glitchy time (248~354Hz injection) is removed.

Images attached to this comment
tatsuki.washimi - 9:44 Tuesday 15 April 2025 (33424) Print this report

Appendix: a case of using only 50-300Hz data in the 50-400Hz injection test

Images attached to this comment